Specification of Sample Injectors
| Orifice | 1.2mm |
| Wetted material | PCTFE, Sapphire |
| Port configuration | 6/8/10 port |
| Port-to-Port volume | 27.5μL |
| Dead volume | 5.4μL |
| Air pressure | 0-1.0Mpa |
| Water pressure | 0-1.6Mpa |
| Origin detection | Auto reset to initial position when powered on (optional and settable) |
| Liquid temperature | 0-150℃ |
| Connection | 1/4-28UNF |
| Replaceable parts | Stator replaceable, sealed rotor irreplaceable |
| Transposition | Random start to any specific port |
| Driver | Non-optional |
| Valve drive | ≤500ms |
| Max. torque | 3N/m |
| Communication | RS232, RS485, CAN |
| Baud rate | RS232/RS485: 9600bps, 9200dps, 38400dps, 57600dps, 115200dps
CAN: 100Kbps, 200Kbps, 500Kbps, 1Mbps |
| Address & Parameter setup | Communication interface |
| Power supply | DC24V/2.0A |
| Max power | 36W |
| Operating temperature | -10℃ to +50℃ |
| Operating humidity | <80% relative humidity, non-condensing |
| Dimension (L*W*H) | 60*51*145.8mm |
| Net weight | 0.73KG |